To Remove And Replace The Gpib Interface Board - Agilent Technologies 54845B Manual

Table of Contents

Advertisement

Chapter 6: Replacing Assemblies

To remove and replace the GPIB interface board

To remove and replace the GPIB interface board

Use this procedure to remove and replace the GPIB interface board. When necessary, refer to
other removal procedures.
Disconnect the power cable and remove the cover.
1
Remove the Torx T15 screw that secures the GPIB board to the rear of the chassis.
2
Pull the board up to disengage it from the motherboard, then lift up and out of the
3
chassis.
CA UT IO N
BE CAREFUL NOT TO SNAG THE ACQUISITION CABLE W11.
To replace the board, reverse the removal procedure.
4
Figure 6-20
Removing the GPIB Interface Board
6–18