N9360A Multi UE Tester W-CDMA Programming Manual
• <int 1> (1 digit) is
measurement range, or
range or ACLR DSB 5MHz is not measured.
• <int 2> (1 digit) is the test result for the pass (0) or fail (1).
• <real 3> (up to 6 letters) is the measured data ranging
from
to
-99.99
0.00
range or ACLR DSB 5MHz is not measured, the value
is returned.
This command is valid only when the W07 option is installed in
the Tester. If the W07 option is not installed in the Tester, an
execution error occurs.
ACLR DSB 10MHz Measurement
TXANalyzer:RESults:ACLR10mhz?
This query command returns a set of test result values, <int
1>, <int 2>, and <real 3>.
• <int 1> (1 digit) is
measurement range, or
range or ACLR DSB 10Hz is not measured.
• <int 2> (1 digit) is the test result for the pass (0) or fail (1).
• <real 3> (up to 6 letters) is the measured data ranging
from
to
-99.99
0.00
range or ACLR DSB 10MHz is not measured, the value
is returned.
-99.99
This command is valid only when the W07 option is installed in
the Tester. If the W07 option is not installed in the Tester, an
execution error occurs.
OBW Measurement
TXANalyzer:RESults:OBW?
This query command returns a set of test result values, <int
1>, <int 2>, and <real 3>. Each value is separated by a
comma.
• <int 1> (1 digit) is
measurement range, or
range or OBW is not measured.
• <int 2> (1 digit) is the test result for the pass (0) or fail (1).
Programming Command Reference
if measurement result is within the
0
if measurement result is out of
1
in dB. If measurement result is out of
if measurement result is within the
0
if measurement result is out of
1
in dB. If measurement result is out of
if measurement result is within the
0
if measurement result is out of
1
5
-99.99
5-71