4
Programming Command Guidelines
Table 4-4 Status Byte Register
Bit
Description
0
0: Measurement data is not ready in the output queue.
1: Measurement data is ready in the output queue.
1
0: Measurement is completed.
1: Measurement is not completed.
2, 3
These bits are always set to 0.
4
0: Measurement data is not ready in the output queue.
1: Measurement data is ready in the output queue.
5
0: Standard Event Status Register Masked by Standard Event Status Enable Register is not set to 1.
1: Standard Event Status Register Masked by Standard Event Status Enable Register is set to 1.
6, 7
These bits are always set to 0.
4-16
Figure 4-6 Status Byte Register
Bit Number
7
6
The Status Byte Register contains the following information:
To query the Status Byte Register, send the command *STB?.
The response is the decimal sum of the bits which are set to
For example, if the bit 5 and bit 4 are set to
of these 2 bits is 48 (32 plus 16) see
value 48 is returned.
Standard Event Status Register
The Standard Event Status Register of the Tester is used to
determine the specific event that set bit 5 in the status byte
register as follows:
N9360A Multi UE Tester W-CDMA Programming Manual
5
4
3
2
1
*
STB?
Figure
0
.
1
, the decimal sum
1
4-4. So the decimal