N9360A Multi UE Tester W-CDMA Programming Manual
This command selects Autoboot setting from NONE, GSM or
WCDMA, and <int> sets the duration time (unit: second) until
start of automatic boot.
• If this is set to None, the Tester does not start automatic
boot.
• If this is set to GSM or WCDMA, the Tester starts automatic
boot with the GSM application or the W-CDMA application,
respectively.
• Preinstalled Option must be installed in the Tester to boot
GSM.
• Preinstalled Option must be installed in the Tester to boot
W-CDMA.
The duration time is within the ranges from
step. At preset, these settings are
1
the Tester does not start automatic boot and the display shows
the [Top Menu] screen.
Programming Command Reference
to
10
60
and 0, which means
NONE
5
seconds in
5-9