Needle Guide Type Preset Selection

A preset is provided in Set Up/Custom Display page 1 to
choose single (SGL) or Multiple (MBX) guidezone angle
capabilities. Available selections for respective probes are
shown in the chart below.
Probe
C358
C364/CBF
C386
C551/CAE
C721
546L
L764/LH
739L
LA39
S316/UC
S317
S611
T739
Table 15–8. Biopsy Guide Attachment Selection
Failure to match the guidezone displayed to the guide may
DANGER
cause the needle to track a path outside the zone.
It is extremely important that when using the adjustable angle
biopsy guides, the angle displayed on the screen matches the
.
angle set on the guide, otherwise the needle will not follow the
displayed guidezone which could result in repeated biopsies or
patient injury.
NOTE: Although the multi-angle guides are compatible with the
Civco Ultrapro and Ultrapro II, it is recommended the
multi-angle guides only be used with the Ultrapro II.
LOGIQ
500 Basic Users Manual
2233658–100 Rev. 0
Depth in cm at Center Channel
Fixed–
Multi–Angle
Angle
(MBX)
SGL
MBX1
MBX2
n/a
4.0
8.0
4.0
n/a
4.0
7.0
4.0
3.0
2.0
n/a
4.0
2.0
2.0
2.0
2.0
1.5
n/a
8.0
n/a
8.0
4.0
n/a
4.0
2.0
2.0
Preparing for a Biopsy
MBX3
6.0
8.0
6.0
8.0
6.0
8.0
5.5
7.0
3.0
4.5
5.5
7.0
4.0
7.0
4.0
6.0
n/a
n/a
n/a
n/a
6.0
8.0
5.5
7.0
4.0
6.0
15–33