Imaging Parameter 10) - GE Logiq 500 User Manual

Hide thumbs Also See for Logiq 500:
Table of Contents

Advertisement

Custom Display
Page 10 of 18 (Imaging Parameter 10)
Image Mode
EC, R
Choose the default imaging mode for this preset.
B
B/B
B/M
B/D
CFM-B
CFM-B/B
CFM-B/M
CFM-B/D
Acoustic Power Knob Effect
Set how the Acoustic Output knob will affect the scan image.
Both B and D—Knob changes power for B/M/D/CFM/PDI all the time regardless of
scan mode.
Only B—Knob changes power for B or B/M mode only. D/CFM/PDI modes cannot be
adjusted by Acoustic power knob. If B or M-Modes are in combination with D/CFM/PDI
they also cannot be adjusted. It is fixed at setup preset value.
Independent B and D—Knob changes power for B/M-Mode while in B/M-Mode. Knob
changes power for D/CFM/PDI mode while in D/CFM/PDI mode.
No Effect—Acoustic output knob has no effect on changing power. Values fixed at set
up preset numbers.
Displayed TI
EC
Choose the type of thermal index display.
TIS
Thermal Index – Soft Tissue
TIC
Thermal Index – Cranial Bone
TIB
Thermal Index – Bone
Image Gamma [%]
EC, UP
Choose the level of gamma correction applied to the image during photography.
Choose from 50 to 200 in 5 digit increments.
Auto Focus Control
EC
Choose to have the focus control follow depth changes as a percent of depth or have
the focus depth be a fixed value.
On
Focus is a percent of the current depth (Focus Pos % Depth preset) until
changed by focus position.
Off
Focus is at a fixed depth until changed by focus position control (Focus Depth
[mm] Preset).
14–30
B-Mode only
Dual B-Mode
B- and M-Modes
B- and Doppler Modes
B-Mode with Color Flow
B-Mode with Color Flow and B-Mode
B- and M-Modes with Color Flow
B- and Doppler Modes with Color Flow
EC
LOGIQ
500 Basic Users Manual
2233658–100 Rev. 0