Page 7 of 18 (Imaging Parameter 7 – Probe Dependent 7)
CFM M Gain (Delta from B)
Choose the default level for M-Mode gain while in CFM.
Level from –31 to 31 in 1 digit increments.
CFM M Wall Filter
EC, UP, P, R
Choose the default level for the M-Mode Wall Filter while in CFM Mode.
Low, M1, M2, M3, M4, or High.
CFM ACE
EC, UP, P, R
Choose the default setting for Adaptive Color Enhancement, which reduces clutter noise
or color flash.
On or Off.
CFM Noise Blanker
EC, UP, P, R
Choose the default setting for Noise Blanker, which reduces random color noise.
OFF, LOW, MID or HIGH.
CFM Persistence
EC, UP, P, R
Choose the default setting for Persistence, which retains highest color pixel value for a
specified period of time or until a higher value is detected.
OFF, SHORT, MID or HIGH.
CFM MR-Flow
EC, UP, P, R
Choose the default setting of the MR-Flow option (combination focus in CFM area).
Limited to LA39, C551 (CAE), C358, E721, 739L and 546L probes only. Must have
ACE-2 option installed.
On or Off.
CFM MR-Flow Combi Number
Set the number of focal zones to be used for the combination focus for MR-Flow option.
2 or 3.
LOGIQ
500 Basic Users Manual
2233658–100 Rev. 0
EC, UP, P, R
EC, UP, P
Custom Display
14–25