Angle (cont'd)
If a fine adjustment of the 2 cursors is needed, press
Measurement to toggle which line cursor is active. Adjust the
active cursor with the Trackball and Zoom Size/Rotation
control.
Press Set to complete the angle measurement.

% stenosis (stenosis ratio)

To calculate percent stenosis in a B-Mode image:
The area of the blood vessel can be measured by the ellipse,
trace, circle, 2 distance or 1 distance methods. See General
Measurements and Calculations for details on the steps to take
when doing basic distance and area measurements. The Set
Up/Preset Program page 3 defines the default method.
Select % Steno in the Calculations Sub-Menu.
VOLUME
ANGLE
Figure 10–25. Cardiac Calculations Sub-Menu
The first cursor is the residual area of the blood vessel and the
second cursor is the lumen area of the blood vessel.
When using a distance measurement to calculate area for
CAUTION
% stenosis, the measurement should always be taken from a
cross sectional view of the vessel.
Do NOT take a distance measurement from a longitudinal view
for an area calculation. This may lead to an inaccurate
assessment of % stenosis.
LOGIQ
500 Basic Users Manual
2233658–100 Rev. 0
Additional Cardiac Calculations
%
Steno
LV
Report
10–23