M–Mode

Optimizing the Timeline

Common Controls
Description
Since M-Mode is basically a single B-Mode scan vector
displayed over time, basic controls that affect the B-Mode
display also affect the M-Mode display.
Acoustic output, TGC, depth and B/M Gain affect both the
M-Mode and B-Mode displays.
.
Scan area size, scan area position and reverse affect B-Mode
only.
If the scan area size is reduced and the position changed, the
M-Mode cursor will follow the position change to stay within the
displayed scan area.
The Dual Format Keys (L–R) work the same as in dual B-Mode,
but display both B-Mode and M-Mode on the left and right side
of the screen.
Accessing/Changing
See the B-Mode section of this chapter for details on these
controls.