Probe Overview

Coupling gels

Do not use unrecommended gels (lubricants). They may
CAUTION
damage the probe and void the warranty.
Applying
In order to assure optimal transmission of energy between the
patient and probe, a conductive gel or couplant must be applied
liberally to the patient where scanning will be performed.
Precautions
Coupling gels should not contain the following ingredients as
they are known to cause probe damage:
S
S
S
S
S
S
S
S
S

Planned Maintenance

The following maintenance schedule is suggested for the
CAUTION
system and probes to ensure optimum operation and safety.
Do the Following
Inspect the Probes
Clean the Probes
Disinfect Probes
Table 15–6. Planned Maintenance Program
15–18
Methanol, ethanol, isopropanol, or any other alcohol-based
product
Mineral oil
Iodine
Lotions
Lanolin
Aloe Vera
Olive Oil
Methyl or Ethyl Parabens (para hydroxybenzoic acid)
Dimethylsilicone
Daily
After Each Use
X
X
X
As Necessary
LOGIQ
500 Basic Users Manual
2233658–100 Rev. 0