Surface Mode (Option); Overview - GE Logiq 500 User Manual

Hide thumbs Also See for Logiq 500:
Table of Contents

Advertisement

Overview

This option creates a projected image different from 3DvieW. It
uses images stored in Cine. The calculation used for the
projection is based on the transmission principle of light using
Volume-Rendering. This is most often used to create a surface
imaging of a fetal face.
This function is available only in single B-Mode.
Refer to 3DvieW Mode section as needed.
This 3D-Surface option is designed for qualitative information
CAUTION
only and has no provision for any measurement or computation
ability.
No specific reconstruction resolution or accuracy is provided or
claimed for this option.
.
Since positional information of the original scan is not available,
CAUTION
all measurement functions cannot be performed on a
3D-Surface image.
NOTE: Consider all of the above cautions when using
3D-Surface images to evaluate a patient. 3D-Surface images
should not be used for quantitative information but a qualitative
supplement to all other diagnostic data collected for that patient.
Refer to the Advanced Reference Manual for further details.