Probe Usage

For details on connecting, activating, deactivating,
disconnecting, transporting and storing the probes, refer
to 3–29 .

Care and Maintenance

Inspecting probes
Inspect the probe's lens, cable, and casing after each use.
Look for any damage that would allow liquid to enter the probe.
.
If any damage is found, do not use the probe until it has been
inspected and repaired/replaced by a GE Service
Representative.
After Each Use
Keep a log of all probe maintenance, along with a picture of any
probe malfunction.
Environmental Requirements
Probes should be operated, stored, or transported within the
parameters outlined below.
Temperature
Humidity
Pressure
Table 15–5. Probe Environmental Requirements
LOGIQ
500 Basic Users Manual
2233658–100 Rev. 0
Operational
Storage
10 _ - 40 _ C
–10 _ - 60 _ C
50 _ - 104 _ F
14 _ - 140 _ F
30-85%
30-90%
non-condensing
non-condensing
700-1060hPa
700-1060hPa
Probe Overview
Transport
–40 _ - 60 _ C
–40 _ - 140 _ F
30-90%
non-condensing
700-1060hPa
15–9