Safety Precautions

Acoustic Output

Controls Affecting Output
The potential for producing mechanical or thermal bioeffects is
influenced by the controls listed below (refer to 2–17 ).
Direct. The Acoustic Output control has the most significant
effect on Acoustic Output.
Indirect. Indirect effects may occur when adjusting the controls
listed on 2–17 .
Always observe the output display for possible effects.
Best practices while scanning
S
Hints
.
NOTE: Refer to the Optimization sections of the Modes chapter
for a complete discussion of each control.
Be sure to have read and understood control explanations for
WARNING
each Mode intended to be used before attempting to adjust the
Acoustic Output control or any control that can affect Acoustic
Output.
Acoustic
Use the minimum necessary output to get the best diagnostic
Output
image or measurement during an examination. Begin the exam
Hazard
with the probe that provides an optimum focal depth and
penetration.
2–16
Raise the Acoustic Output only after attempting image
optimization with controls that have no affect on Acoustic
Output, such as Gain and TGC.
LOGIQ
500 Basic Users Manual
2233658–100 Rev. 0