Overview

The Set Up/Utility Menu, option #04 allows the user to back-up
(save) all user preset data to MOD. In the event of a system
upgrade or catastrophic hard drive failure, the user application
presets can be easily loaded back to the system if they have
been previously saved to MOD.
Once presets have been backed up, the user data back-up
should be performed periodically to update any changes that
have been made to the user preset parameters.

Saving Presets

Before saving presets, the MOD must first be initialized
(formatted) prior to saving user application preset data.
S
S
S
.
The system then takes the time to format the MOD (approx. 12
minutes).
NOTE: If presets have already been saved on the MOD, the
MOD should be initialized again to ensure that old preset data
has been deleted from the MOD.
To save presets:
S
S
LOGIQ
500 Basic Users Manual
2233658–100 Rev. 0

User Data Back-up

Select Set Up/Utility Menu.
Select #09 Media Initialize and press Return.
Insert a MOD into the drive and press Return.
Enter #04 User Data Backup from the Utility Menu and
press Return.
Enter # 1 (Save) and press Return.