ET (Ejection Time)

Ejection time for the left ventricle is the time during which the
aortic valve is open. It is measured on a Doppler Spectrum
display. The time between the two cursors is ejection time (ET).
Figure 10–28. Ejection Time for Aortic Valve
Select ET from of the Calculation Sub-Menu. Perform a time
interval measurement as described on 7–18 . Set the first
cursor at the opening of the aortic valve and the second at the
closing of the aortic valve.
Figure 10–29. Cardiac Calculation Sub-Menu
LOGIQ
500 Basic Users Manual
2233658–100 Rev. 0
Additional Cardiac Calculations
PHT
ET
Max PG
MVA
Mean PG
TRACE
AUTO
LV
Report
10–25