Warning Message Description - GE Logiq 500 User Manual

Hide thumbs Also See for Logiq 500:
Table of Contents

Advertisement

Operation Guide Message Description (cont'd)
Category
Preset & Setup
Data is saved.
Function
Input value by Rocker
Button.
Input character strings.
Select an initial data.
'TRACKBALL' and 'SET'
are available.
Recommended value are
copyed.
Table 16–5. Operation Guide Message Description (cont'd)

Warning Message Description

Category
Temperature Safety
WARNING:
Temperature error.
WARNING: Cooling Fan
WARNING
Failure.
WARNING: Cooling Fan
Failure. Temperature
error.
WARNING: HV
Abnormal.
WARNING: Probe
Temperature error.
WARNING: NOW
START THE POWER
OFF PROCESS.
Table 16–6. Warning Message Description
LOGIQ
500 Basic Users Manual
2233658–100 Rev. 0
Problem/
Possible
Message
Cause
The system completed
saving the data that was
requested.
Operator can use rocker
switch to input the value.
Operator needs to input
some character strings.
Operator needs to select
initial data.
Operator can use the
Trackball and Set key to
operate the function.
The system completed
copying the data that
was requested.
Problem/
Possible
Message
Cause
System internal
temperature is at a
maximum, but not
caused by fan failure.
System fan has stopped
working properly.
System internal
temperature is at a
maximum, caused by the
fan not working properly.
The High Voltage is
abnormal.
The probe temperature
is at a maximum.
The system shutdown
process has begun.
Troubleshooting
Possible
Corrective Action
NA
Use the Ellipse rocker
switch.
Type in characters.
Input proper data.
NA
NA
Possible
Corrective Action
Reduce room
temperature.
Call service.
Reduce room
temperature; call service.
Call service.
Call service.
Call service.
16–33

Advertisement

Table of Contents
loading

Table of Contents