Peripheral/Accessory Connection - GE Logiq 500 User Manual

Hide thumbs Also See for Logiq 500:
Table of Contents

Advertisement

Console Overview

Peripheral/Accessory Connection

Peripheral/Accessory Connector Panel
LOGIQ
connected using the rear connector panel located behind the
rear door. Only the B/W Page Printer (UP-890) can be
connected to the front accessory panel.
Located on the rear panel are video input and output, audio
input and output, camera expose, foot switch, power and
control connections for VCR, printer, MIC and service tools.
See Figure 3–5.
Each outer (case) ground line of peripheral/accessory
CAUTION
connectors are Earth Grounded.
Signal ground lines are Not Isolated, except the Service Port.
All of the signal lines (including the signal ground) of the
Service Port are Isolated.
Use only approved probes, peripherals or accessories.
CAUTION
3–8
500 peripherals and accessories can be properly
Service
Figure 3–4. Service Port
LOGIQ
500 Basic Users Manual