General Instructions; Erasing Measurements - GE Logiq 500 User Manual

Hide thumbs Also See for Logiq 500:
Table of Contents

Advertisement

Introduction

General Instructions

Prior to making measurements, Freeze should be pressed to
stop the acquisition of the image data.
For some types of measurements, press the Measurement key
to toggle between active cursors for fine adjustment prior to
completing the measurements.
Pressing Clear prior to completing the measurement sequence
erases the active measuring marker and the current data
measured.
Pressing Clear after the sequence is complete erases all data
that has been measured to this point, but not data entered on to
report pages.
Any measurement can be repeated by selecting that
measurement again from the Sub-Menu.

Erasing Measurements

The following actions erase measurements from the system's
memory:
S
S
S
S
S
"Measurement Clear Operation" is a parameter found in the Set
Up/Preset Program page 3. This parameter can be set to have
only measurements erased with the Clear key or
measurements and comments.
7–6
Rotating the Cine Scroll knob erases measurements.
Pressing New Patient erases all measurements and
calculations on the display and clears the report pages.
Unfreezing the image erases measurements from the
display only, not from the report pages.
Adding a new measurement that exceeds the maximum
number of allowable measurements erases the first or
oldest measurement.
Pressing Clear erases all completed measurements and
calculations on the display. Measurements and
calculations, however, remain on the report pages.
LOGIQ
500 Basic Users Manual
2233658–100 Rev. 0

Advertisement

Table of Contents
loading

Table of Contents