Imaging Parameter 4 - Probe Dependent 4) - GE Logiq 500 User Manual

Hide thumbs Also See for Logiq 500:
Table of Contents

Advertisement

Custom Display
Page 4 of 18 (Imaging Parameter 4 – Probe Dependent 4)
PWD/CFM/PDI Penetration
Choose to have the default for penetration, using the lower probe frequency, for pulsed
Doppler, color flow mapping and power Doppler imaging modes on or off.
Off
On
PWD/CFM/PDI Slant Auto Start
Choose the default position of the slant scan window.
Off
Other-Side
Marker-Side
PWD Gain
EC, UP, P, R
Choose the default gain value for pulsed Doppler Mode.
Value from 0 to 32 in 2 digit increments.
PWD Dynamic Range
EC, UP, P, R
Choose the default dynamic range value for pulsed Doppler Mode.
Value from 18 to 48 in 6 digit increments.
PWD Sample Volume [mm]
Choose the default pulsed Doppler sample volume cursor size (in millimeters).
1, 2, 3, 4, 5, 6, 7, 8, 9, 10, 12, 14 or 16.
PWD Gray Scale Map
EC, UP, P, R
Choose the pulsed Doppler Mode gray scale default map.
1, 2, 3, 4, 5, 6, 7, 8, 9 or 10.
PWD Rejection
EC, UP, P, R
Choose the pulsed Doppler Mode rejection default level.
Level from 0 to 40 in 2 digit increments.
PWD Velocity [cm/s]
EC, UP, P, R
Choose the pulsed Doppler Mode velocity scale default value.
Value from 10 to 1000 in 10 digit increments.
PWD Wall Filter [cm/s]
Choose the pulsed Doppler Mode wall filter default value.
0.3, 0.5, 0.7,1.0, 1.5, 2.0, 3.0, 5.0, 7.0, 10, 15, 20, 30 or 50.
14–20
EC, UP, P, R
Use higher probe frequency.
Use lower probe frequency.
EC, UP, P, R
Vertical (no slant)
Slanted away from the probe orientation mark
Slanted toward the probe orientation mark
EC, UP, P, R
EC, UP, P, R
LOGIQ
500 Basic Users Manual
2233658–100 Rev. 0