ECG Option
ECG Sub-Menu Page 1
Figure 10–38. ECG Sub-Menu Page 1
Single
Description
Updates the B-Mode image at the R1 time.
Each time the R1 point is reached on the
ECG, PCG, or AUX Waveform, the
B-Mode image changes. In between R1
trigger points, the B-Mode image is frozen.
The display shows the delay value as S1.
Benefits
Allows for the acquisition of an image at a
specific point in the Cardiac Cycle.
Values
If Single is highlighted, the one-trigger
function is on. If Single is not highlighted,
the function is off.
NOTE: The Trackball can be used to
change the value of the active delay.
Affects on
If Dual was highlighted when single is
pressed, the Dual function is turned off
Other Controls
and the Single function turned on.
10–34
B
Preset
Set Up
Single
Sync.
R
Selectn
Delay
Dual
Ref.
ECG
Scan
Wave
ECG
PCG
Wave
Aux
Wave
Dual
Updates the B-Mode image at the R1 and
R2 times. Each time the R1 and R2 points
are reached on the ECG, PCG, or Aux
waveform, the B-Mode image changes. In
between each trigger time, the B-Mode
image is frozen.
NOTE: R
on the ECG will be marked with
2
a vertical dotted line. The display shows
the delays as S1 and S2.
Allows for the acquisition of an image at
two points in the Cardiac Cycle.
If Dual is highlighted, the two-trigger
function is on. If Dual is not highlighted,
the two-trigger function is off.
NOTE: The Trackball can be used to
change the value of the active delay.
If Single is highlighted when Dual is
pressed, the single function is turned off
and the Dual function turned on.