Description Of The Innova 2000 Imaging System; Description Of The Tv Monitor & Rad Shield Suspensions; Description Of The Dl - GE Innova 2000 Preinstallation Manual

Cardiovascular imaging system
Table of Contents

Advertisement

GE Medical Systems
REV 5
2-5

Description of the Innova 2000 Imaging System

An Innova 2000 System is managed and controlled by a System including:
D Atlas cabinet (inside which are located the RTAC, the DL, the modem, the HUB, the KVM, the
4 KVA PDU),
D Innova Console with a plasma screen mounted on a wall support or pedestal,
D 22 cm Revolution Digital Detector mounted on Innova LC positioner.
D An external Digital Detector chiller is mandatory.
The distance cannot exceed 3 m (1 floor) below the detector
CAUTION
D 53 cm (21") 21" Hi–Brite monitor,
D 43 cm (17") 17" Hi–Brite monitor (optional).
2-6
Description of the TV monitor & Rad Shield suspensions
The common type of this suspension is an XT inboard monitor bridge.
A monitor frame support receiving 3 CRT monitors (2x Hi–Brite monitors and one monitoring).
A monitor frame support receiving 4 CRT monitors (2x Hi–Brite monitors and 2 monitorings) (to be
defined later).
A monitor frame support receiving 3 or 4 LCD monitors.
Only for Control Area a Gyro TVR wall mount can supports the 2 Hi–Brite monitors 53 cm or 43 cm
(21" or 17").
Rad Shield
A counterbalanced radiation shield suspension is supplied as an option. A high intensity lamp can
also be added.
2-7

Description of the DL

An Innova 2000 System uses the DL Digital Imaging system.
A complete DL System contains the following:
D included in the Atlas Cabinet:
– DL Computer,
– modem,
– Ethernet HUB,
– KVM,
– Video Splitter,
Innova 2000 Cardiovascular Imaging System
11
Pre–Installation Manual
pim 2337741–100