Activation Of Alarms After Breath Detection; Intelligent Alarm Behavior; Combined Alarms; Suppressed Alarms - Dräger Perseus A500 Instructions For Use Manual

Anesthesia workstation
Hide thumbs Also See for Perseus A500:
Table of Contents

Advertisement

Alarms

Activation of alarms after breath detection

All O
alarms, CO
alarms, N
2
2
anesthetic gas alarms are only active once 2
breaths have already been detected.
If no breaths have yet been detected when leaving
the modes Standby or Pause, the message
Waiting for respiratory phases is displayed in
the CO
waveform.
2

Intelligent alarm behavior

Combined alarms

If multiple alarms occurring at the same time are
caused by the same problem, they are combined
into one alarm.
Problem
Several causes of apnea are
present.
Faults in multiple components.
This causes failure of a system
function.

Suppressed alarms

Some low-priority alarms indicate a fault in a
measurement function. If this measurement
function monitors physiological parameters, alarms
based on these parameters will not be generated.
Example:
Fault
CO
measurement is faulty
2
124
O alarms, and
2
Alarms occurring at the same
time
Apnea (no flow)
Apnea (no pressure)
)
Apnea (no CO
2
Example:
Insp. press. sensor failure
Exp. press. sensor failure
Displayed alarm
Sample line occluded
Once 2 breaths have been detected, the message
disappears and only then the O
alarms, N
O alarms, and the anesthetic gas
2
alarms are active.
Combined alarm
Apnea
Ventilator failure
Non-generated alarm
Apnea (no CO
Instructions for use Perseus A500 SW 1.1n
alarms, CO
2
2
)
2

Advertisement

Table of Contents
loading

Table of Contents