Measuring CO

2
Displaying CO
Waveform
2
Press »Waves« key.

Setting Alarm Limits for etCO

Set alarm limits are maintained even when the ventilator
is switched off.
Evita 2 dura will generate an alarm if the value for etCO
remains continuously outside the set alarm limits for
longer than 15 seconds.
Adjustment
range
1 to 100 mmHg
0.1 to 15 kPa
_
etCO2
0 to 99 mmHg
0 to 14.9 kPa
Press »Alarms« menu key.
The »Limits« menu page will appear
Set alarm limit with the dial knob and press to
confirm. The new alarm limit will now be in effect.
––––––––––––––––
*
see also Theory of Operation, "Physiological Capnogram, p. 27
mmHg
80
60
40
20
0
0
2
2
Factory-set
start-up defaults
30 mmHg
60 mmHg
Measuring CO
Displaying CO
Setting Alarm Limits for etCO
CO
2
2
4
s
2
Waveform
2
2
13