User Adjustable Alarm Limits - Dräger NAD Operator's Instruction Manual

Omicron monitor
Table of Contents

Advertisement

6
Specifications

User Adjustable Alarm Limits

The following displays the user adjustable alarm limits specification
information:
Table 1. User Adjustable Alarm Limits
Alarm Limit Type
Pressure High Limit
Pressure Low Limit
O
2
O
2
Total Respiratory Rate High Limit
Tidal Volume Low Limit
Tidal Volume High Limit
Minute Volume Low Limit
Minute Volume High Limit
6-6
High Limit
Low Limit
RETURN TO CD-ROM TABLE OF CONTENTS
Alarm Limit Value
+11 to +120 cm H
(Resolution 1 cmH
+5 to +114 cm H
(Resolution 1 cmH
25% to 99%
100% = High Alarm Disabled
(Resolution 1%)
18% to 95% (Resolution 1%)
5 to 250 BPM (Resolution 5 BPM)
50 to 2400 mL (Resolution 50 mL)
150 to 2500 mL (Resolution 50 mL)
0 to 79 L (Resolution 1 L)
1 to 80 L (Resolution 1 L)
O
2
O)
2
O
2
O)
2