Example for sounding the alarm tone when
several alarms are present
Priority of the existing alarm
(!!!) WARNING
(!!!) WARNING
(!!) CAUTION
(!!) CAUTION

Suppressing the alarm tone

The alarm tone can be suppressed for a maximum
of 2 minutes.
B
 Press the
(A) key.
The LED in the key (A) lights up.
In the status bar (B), the symbol
remaining time for the alarm tone suppression are
displayed.
During the alarm tone suppression, only the new
alarms are acoustically signaled whose alarm
priority or internal priority number is higher than
the suppressed alarm, see chapter ''Alarm –
Cause – Remedy'' on page 152.
Instructions for use Fabius Tiro SW 3.n
Priority of the new alarm
(!!!) WARNING
(!!) CAUTION
(!!) CAUTION
(!!!) WARNING
Reactivating the suppressed alarm tone
 Press the
The LED in the key (A) goes out.

Adjusting the alarm limits

A
If an alarm is triggered due to a value falling below
or exceeding the alarm limit, it might be necessary
to adjust the standard alarm limits. More
information for setting the alarm limits can be
found in chapter ''Changing the alarm limits'' on
page 134.
Setting the alarm limits in current ventilation mode:
 Press the
More information can be found in chapter
and the
''Monitoring'' on page 116.
Reaction from Fabius
– Alarm tone sequence starts
from the beginning.
– Alarm tone sequence for the
existing alarm is not inter-
rupted.
– No acoustic alarm signal for
the new alarm
– Alarm tone sequence starts
from the beginning.
– Alarm tone sequence for the
new alarm is started.
(A) key again.
key.
Alarms
115