Checking Height Adjustment (Optional); Switching On And Activating The Self-Test - Dräger Babytherm 8004 Instructions For Use Manual

Hide thumbs Also See for Babytherm 8004:
Table of Contents

Advertisement

Testing Readiness for Operation

Checking height adjustment (optional)

Test height adjustment system if fitted:
1
1 Briefly press right pedal.
The bed is raised.
2 Briefly press left pedal.
The bed is lowered.
 Remember the maximum load for height
adjustment is 50 kg.
 Hoses and cables should be long enough to
ensure a secure connection even in the top or
bottom height adjustment position.
 Do not place any objects in the raising/lowering
path.
 Adjust to a comfortable working height.

Switching on and activating the self-test

 Press the On/Off button until it engages = ON.
Babytherm now runs a self-test to check
important functions.
40
2
3 The following message appears on the display:
All displays on, horn on
All displays are switched on for approx.
2 seconds:
All LEDs are lit; the numeric displays show
88.88 and a continuous tone is emitted.
3 The following message appears on the display:
All displays off, soft alarm on
All displays are dark for about 2 seconds and
the alarm sounds.
4 The green LED is lit.
5 After about 2 seconds, the unit displays the
radiant warmer mode:
man. (manual mode) – no skin temperature
sensor in place
or
Skin (baby control mode) – yellow skin
temperature sensor for core temperature in
place.
The corresponding LED flashes.
6 The preset desired values are displayed.
3 If the text display reads Battery discharged
7 and the yellow LED is lit, the battery is being
charged for the power failure alarm. It takes
about 30 minutes to charge the battery.
3
10
5
1
6
skin C
manual
mattress. C
skin
man.
5
Instructions for use Babytherm 8004/8010
Babytherm 8010
7
15 min
C
8
< 36 C
6
4

Advertisement

Table of Contents
loading

This manual is also suitable for:

Babytherm 8010

Table of Contents