Summary of Contents for Keysight Technologies N9040B

  • Page 1 Keysight Technologies N9040B UXA Signal Analyzer Option B2X, 255 MHz Analysis Bandwidth Notice: This document contains references to Agilent. Please note that Agilent’s Test and Measurement business has become Keysight Technologies. For more information, go to www.keysight.com Installation Note...
  • Page 2 The information contained in this document is subject to change without notice. Keysight Technologies makes no warranty of any kind with regard to this material, including but not limited to, the implied warranties of merchantability and fitness for a particular purpose.
  • Page 3 Option B2X, 255 MHz Analysis Bandwidth Option B2X, 255 MHz Analysis Bandwidth Products Affected UXA N9040B Serial Numbers: To Be Performed By: (X) Keysight Service Center ( ) Personnel Qualified by Keysight ( ) Customer Estimated Installation Time: 2.0 Hours Estimated Adjustment Time: 6.0 Hours...
  • Page 4: Tools Required

    Option B2X, 255 MHz Analysis Bandwidth Contents Quantity Description Keysight Part Number Installation Note This note Option Upgrade Entitlement Certificate 5964-5178 Wideband Analog IF assembly N9020-60043 Wideband Digital IF assembly N9020-60311 Flex Circuit, WB DIF to Flex Circuit Interconnect Bracket N9040-60045 Flex Circuit, WB AIF to Flex Circuit Interconnect Bracket N9040-60042...
  • Page 5: Initial Instrument Functionality Check

    Option B2X, 255 MHz Analysis Bandwidth Initial Instrument Functionality Check 1. Power on the instrument and allow the instrument to boot up, run the alignments and display the measurement screen. The instrument will probably display a spectrum analyzer screen and you will see the instrument sweeping.
  • Page 6 Option B2X, 255 MHz Analysis Bandwidth Wideband Analog IF Assembly Hardware Installation 1. Power down the instrument, wait until the standby light comes on, and remove the power cord. 2. Remove the instrument side strap handles, feet, outer case, top brace front, top brace rear, and remove the rear panel (be careful to not damage the cables), and remove the rear brace.
  • Page 7 Option B2X, 255 MHz Analysis Bandwidth 8. Locate the W45 cable with color code 2 on one end and 17 on the other end. Attach color code 2 end to the Wideband Analog IF at J2. Route the cable as shown in Figure 2, the other end of the cable will be routed later to the A3 Digital IF.
  • Page 8 Option B2X, 255 MHz Analysis Bandwidth 13.See Figure 3 that shows where the color code 15 end of W40 connects to the A3J15, Digital IF assembly. Figure 3 DIF Cables Showing W40 Connection to J15 Installation Note N9040-90017...
  • Page 9 Option B2X, 255 MHz Analysis Bandwidth Wideband Digital IF Assembly Hardware Installation 14.See Figure 4. Locate the Wideband Digital IF assembly in the kit. Install the assembly into the top card slot as shown. Figure 4 Wideband Digital I.F. Installation 15.Locate the N9040-60045 Flex Circuit cable in the kit and connect one end to the Wideband...
  • Page 10 Option B2X, 255 MHz Analysis Bandwidth 17.Locate the spring clip in the kit. Install the spring clip to secure the flex circuit connection on the Wideband Digital IF assembly. 18.Re-install the Rear Brace and the Rear Panel. An additional flat head screw, 0515-1946 required to secure the new Wideband Digital IF assembly to the rear panel.
  • Page 11: License Installation Procedure Over Usb

    Option B2X, 255 MHz Analysis Bandwidth License Installation Procedure over USB 1. Locate the Option Upgrade Entitlement Certificate (5964-5178) from the kit. 2. Redeem the Option Upgrade Entitlement Certificate by following the instructions on the Certificate. 3. After redeeming your Option Upgrade Entitlement Certificate you will receive an email with an attached License File.
  • Page 12 2. Press System, Show System to display a list of installed options. 3. Verify that the installed options list contains the newly installed N9040B-B2X. You will see Option DP4 appear also, and this is the extra capture memory the new Wideband Digital IF assembly provides.
  • Page 13 Option B2X, 255 MHz Analysis Bandwidth Instrument Software Installation This upgrade requires instrument software version A.17.07 or later. To verify the current version, press System, Show System and look for the "Instrument S/W Revision". Even if the instrument software is the latest, reinstalling the software will reprogram the FPGAs in the new Wideband Digital IF assemblies.
  • Page 14 Performance Testing Required Verification Test Name Perform all tests For assistance, contact your nearest Keysight Technologies Sales and Service Office. To find your local Keysight office access the following URL, or if in the United States, call the following telephone number: http://www.keysight.com/find/assist...