11. Use the finger-tight PEEK nut and ferrule to attach the free end of the liner tube to the rear
of the probe-tip support.
12. Firmly tighten the finger-tight PEEK nut and ferrule to hold the capillary in place.
13. Carefully pass the probe-tip assembly over the exposed capillary.
14. Using the 4-mm wrench, screw on the probe tip until the end of the capillary protrudes 0.5
to 1.0 mm from the probe tip.
15. Insert the reference-sprayer assembly into its support assembly with the sprayer tip
pointing upward.
16. Rotate the reference-sprayer assembly 90º counterclockwise to lock it in place.
17. Close the source enclosure.
7.21 Replacing the LockSpray II reference probe
The LockSpray II reference probe differs from the original LockSpray reference probe in that it
comes as one complete unit, pre-built to length. The capillary end connects to the reference
sprayer assembly within the LockSpray II source housing.
Important:
There is no LockSpray II retrofit option for the LockSpray source, but the front-end
housings are interchangeable on the instrument.
Replace the LockSpray II reference probe if it becomes blocked and you cannot clear it, or if it
becomes contaminated or damaged.
Notice:
To prevent pressure from ejecting the capillary from the reference-
sprayer assembly, firmly tighten the finger-tight PEEK nut and ferrule.