5. Using a long, finger-tight fitting, connect orange, 1/16-inch, 680-mm PEEK tubing from port
3 of the LockSpray selector valve through tubing guide C and into reservoir bottle C.
Requirement:
As the tubing emerges from the tubing guide, thread the long, finger-tight
fitting over it, push the tubing through the left-hand hole to the bottom of the reservoir
bottle, and finger-tighten the fitting.
6. Using a long, finger-tight fitting, connect orange, 1/16-inch, 1000-mm PEEK tubing from
port 6 of the LockSpray selector valve to the wash bottle.
Requirement:
Push the tubing to the bottom of the wash solution and secure the tubing
to prevent it from floating to the surface during use.
7. Using a long, finger-tight fitting, connect 1/16-inch, 1000-mm, natural-color PEEK tubing to
port 5 of the LockSpray selector valve and thread the tubing into the waste port.
Figure 7–133: Connecting the LockSpray selector valve and the waste port
A
A
B
B
C
C
Wat ers
Waste port
Tip:
The liquid waste system collects waste without requiring a fitted connection. Waste
drains through the connector at the base of the instrument and into a waste bottle (see
Connecting liquid waste lines (Page
8. If the system lacks an optional flow sensor and grounded union, skip to the final step in this
procedure to close the access doors to the IntelliStart Fluidics system.
9. Using a long, finger-tight PEEK nut and 1/32-inch PEEK ferrule, connect the red, 1/32-inch,
200-mm PEEK tubing from port 4 of the LockSpray selector valve to the left-hand side of
the flow sensor.
10. Connect the 1/32-inch, 6-40 Valco compression fitting assembly.