Figure 7–120: Probe and nebulizer gas seals
Probe adjuster nebulizer gas seal
Probe adjuster assembly probe seal
3. Use the O-ring removal kit to carefully remove the source enclosure, nebulizer gas, and
desolvation gas seals from the source enclosure.
Figure 7–121: Removing the source enclosure, nebulizer gas, and desolvation gas
seals
Desolvation gas seal
Source enclosure seal
Nebulizer gas seal
March 31, 2021, 715004496 Ver. 04 (previously released as Rev. D)