Table 4–2: Tubing dimensions (sample delivery system)
Valve/port
Connection
Sample/1
Waste
reservoir
Sample/2
Wash reservoir 0.020
Sample/3
Diverter valve
Sample/4
Reservoir
Bottle A
Sample/5
Reservoir
Bottle B
Sample/6
Reservoir
Bottle C
Sample/7
Analyte pump
Diverter/4
Waste
Required materials
• Chemical-resistant, powder-free gloves
• Needle-nose pliers
• The Xevo G2-XS QTof Fluidics Tubing and Fitting Kit
Tip:
This kit contains components for both the sample and NanoSpray system's plumbing.
To plumb the analyte system:
Warning:
To avoid personal contamination with biologically hazardous, toxic, and corrosive
materials, wear chemical-resistant, powder-free gloves when performing this procedure.
Requirement:
You must finger-tighten all PEEK fittings.
1. Open the access door to the IntelliStart fluidics.
2. Using the stainless steel tubing, connect the sample pump to port 7 of the sample selector
valve.
Requirement:
tighten the fittings appropriately.
March 31, 2021, 715004496 Ver. 04 (previously released as Rev. D)
ID (inch)
0.040
0.005
0.020
0.020
0.020
0.040
0.040
The steel tubing requires high-pressure seals at each end. You must
Page 61
OD (inch)
Color
1/16
Natural
1/16
Orange
1/16
Red
1/16
Orange
1/16
Orange
1/16
Orange
1/16
N/A
1/16
Natural
Length (mm)
1,000
1,000
200
680
680
680
500
1,000