Waters Xevo G2-XS QTof Overview And Maintenance Manual page 189

Hide thumbs Also See for Xevo G2-XS QTof:
Table of Contents

Advertisement

Figure 7–103: Removing the NanoLockSpray reference probe's fixing screws
Fixing screw
NanoLockSpray reference probe
Warning:
reference-probe tip is an exposed, fused-silica TaperTip emitter that is sharp and
fragile.
7. Remove the NanoLockSpray reference probe from the probe adjuster assembly.
8. Unscrew the TaperTip-emitter PEEK coupler, and remove the TaperTip emitter from the
union.
Figure 7–104: Removing the TaperTip emitter
TaperTip-emitter PEEK coupler
Capillary PEEK coupler
Capillary with PEEK protective sleeve
March 31, 2021, 715004496 Ver. 04 (previously released as Rev. D)
To avoid puncture wounds, handle the probe with care. The
Page 189