4.3.2 Moving the sprayer platform into the source
To move the sprayer platform into the source:
Warning:
is in place over the sprayer.
1. Confirm that the clear sprayer shield is in place and secured (see
NanoLockSpray source (Page
2. Confirm that the sprayer's safety cover is installed.
3. Pull out the side thumbscrew on the front of the sprayer platform, and push the sprayer
platform into the source.
4. Release the side thumbscrew to lock the platform in position.
5. Tighten the front thumbscrew, securing the adjuster assembly rigidly to the source.
4.4 Adjusting the sprayer tip position
To adjust the tip position:
1. Use the X-, Y-, and Z-position adjuster controls on the adjuster assembly to move the
sample-sprayer tip close to the sampling cone and baffle (see
57)).
2. Fine-tune the position of the sprayer while acquiring a spectrum of a standard compound.
See also:
NanoLockSpray source.
Tips:
• If you observe an electrical discharge between the sprayer tip and baffle, move the tip
farther from the baffle or reduce the capillary voltage. The capillary voltage must be
high enough to maintain a good spray.
• Small adjustments to the sprayer position can make large differences to the source
sensitivity.
4.5 Setting up the camera
Requirement:
available.
To avoid harmless, static-like electrical shock, ensure that the safety cover
53)).
The instrument's online Help file for information about optimizing the
The NanoLockSpray source must be fitted to the instrument for this option to be
March 31, 2021, 715004496 Ver. 04 (previously released as Rev. D)
Page 57
Overview of the
Setting up the camera (Page