Waters Xevo G2-XS QTof Overview And Maintenance Manual page 195

Hide thumbs Also See for Xevo G2-XS QTof:
Table of Contents

Advertisement

a. Use the wash bottle containing 1:1 methanol/water to rinse the union over the large
beaker.
b. Clear and blow dry the union with inert, oil-free gas.
7.25.1.1 Installing the new TaperTip emitter and capillary
Required materials
• Chemical-resistant, powder-free gloves
• Combined 2.5-mm hex wrench and cone extraction tool
• TaperTip emitter
• A 375-mm length of 75-µm PEEK capillary tubing, sleeved for protection by 360-µm ID PEEK
tubing
• Stainless steel wire plug
• F-130 finger-tight nut
• 1.5-mm hex wrench
To install the new TaperTip emitter and capillary:
1. If you did not disturb the preset position of the union when you removed the TaperTip
emitter and capillary, go to step 2. If you disturbed the position:
• Loosen the set screw using the 1.5-mm hex wrench.
• Reposition the union so the surface is level with the bottom of the body holder.
• Tighten the set screw using the 1.5-mm hex wrench.
March 31, 2021, 715004496 Ver. 04 (previously released as Rev. D)
Page 195