Maintaining The Roughing Pump - Waters Xevo G2-XS QTof Overview And Maintenance Manual

Hide thumbs Also See for Xevo G2-XS QTof:
Table of Contents

Advertisement

Software
MassLynx
UNIFI
2. Pull the source enclosure release (located at the bottom, right-hand side) outward and
swing open the enclosure.
Warning:
corrosive materials, wear chemical-resistant, powder-free gloves when performing this
procedure.
3. Unscrew and remove the nitrogen exhaust trap bottle from the cap and associated fittings.
Warning:
corrosive materials, dispose of all waste materials according to local environmental
regulations.
4. Dispose of the waste liquid in accordance with local environmental regulations.
5. Fit and tighten the nitrogen exhaust trap bottle to the cap.
6. Secure the nitrogen exhaust trap bottle in the upright position.
7. Close the source enclosure.
Tip:
An automatic pressure test is performed.
8. To start the LC flow, follow the action specified below, according to your software:
Software
MassLynx
UNIFI

7.11 Maintaining the roughing pump

You can use the mass spectrometer with either a Leybold oil-filled roughing pump or an Edwards
XDS46i oil-free roughing pump.
To avoid personal contamination with biologically hazardous, toxic, and
To avoid spreading contamination with biologically hazardous, toxic, and
March 31, 2021, 715004496 Ver. 04 (previously released as Rev. D)
Action
In the instrument console, click Stop Flow
.
On the System Console tool bar, click Stop
Flow
.
Action
In the instrument console, click Start Flow
.
On the System Console tool bar, click Start
Flow
.
Page 102