Figure 7–48: Fitting the seal breaker and locator tool
StepWave assembly
Adapter housing
Inverted seal breaker and locator tool
8. Push firmly on the seal breaker and locator tool until the tool's face contacts the adapter
housing.
Rationale:
9. Remove the seal breaker and locator tool.
7.16.8 Fitting the ion block support to the source
Required materials
• Chemical-resistant, powder-free gloves
• 3-mm hex wrench
• New seals and O-rings
To fit the PEEK ion block support to the source:
Warning:
To avoid personal contamination with biologically hazardous, toxic, and corrosive
materials, wear chemical-resistant, powder-free gloves when performing this procedure.
1. Ensure that the grooves for the PEEK ion block support O-rings are free from dirt and
debris.
This fully locates the StepWave assembly in the adapter housing.
March 31, 2021, 715004496 Ver. 04 (previously released as Rev. D)
Page 145