Waters Xevo G2-XS QTof Overview And Maintenance Manual page 185

Hide thumbs Also See for Xevo G2-XS QTof:
Table of Contents

Advertisement

Figure 7–97: Securing the reference sprayer capillary
Reference sprayer assembly
Finger-tight fitting
3. Using the 4-mm open-end wrench, adjust the reference sprayer to ensure the correct
capillary protrusion as required.
4. Insert the reference sprayer assembly into its support assembly with the reference sprayer
tip pointing upward.
Tip:
Align the stem on the reference sprayer assembly with the guide hole in the reference
sprayer support assembly.
Figure 7–98: Inserting the reference sprayer assembly into its support assembly
Reference sprayer support assembly
Locating notch on reference sprayer
Reference sprayer assembly
Reference sprayer support assembly guide hole
March 31, 2021, 715004496 Ver. 04 (previously released as Rev. D)
Page 185

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents