Waters Xevo G2-XS QTof Overview And Maintenance Manual page 196

Hide thumbs Also See for Xevo G2-XS QTof:
Table of Contents

Advertisement

Figure 7–108: ionKey reference probe
IonKey reference probe
Set screw
Body holder
Union
Locating tab
2. Thread the sleeved PEEK capillary through the body of the reference probe.
3. Slide a PEEK coupler over the end of the 75-µm PEEK capillary.
4. Locate the PEEK coupler in the top end of the union and carefully slide the 75-µm PEEK
capillary into the union until it butts against the internal surface of the union.
Rationale:
Doing so ensures a minimum dead volume.
5. Tighten the PEEK coupler to hold the capillary in place.
Warning:
reference-probe tip is an exposed, fused-silica TaperTip emitter that is sharp and
fragile.
March 31, 2021, 715004496 Ver. 04 (previously released as Rev. D)
To avoid puncture wounds, handle the probe with care. The
Page 196