Figure 7–107: Removing the TaperTip emitter
TaperTip-emitter PEEK coupler
Capillary PEEK coupler
Capillary with PEEK protective sleeve
Capillary
Union
TaperTip emitter
TaperTip emitter outlet
8. Unscrew the capillary PEEK coupler and remove the capillary from the union.
9. Where appropriate, remove the protective PEEK sleeve from the capillary for reuse.
Warning:
To avoid spreading contamination with biologically hazardous, toxic, and
corrosive materials, dispose of all waste materials according to local environmental
regulations.
10. Dispose of the capillary and TaperTip, observing local environmental regulations.
11. Using the 1.5-mm hex wrench, unscrew the union retaining screw and remove the union.
12. Remove any remaining shards of glass or other contaminants by cleaning the union as
follows:
March 31, 2021, 715004496 Ver. 04 (previously released as Rev. D)
Page 194