Installing The Unispray Source; Installing The Unispray Lockspray Source - Waters Xevo G2-XS QTof Overview And Maintenance Manual

Hide thumbs Also See for Xevo G2-XS QTof:
Table of Contents

Advertisement

6.1 Installing the UniSpray source

6.1.1 Installing the UniSpray LockSpray source

Required materials
• Chemical-resistant, powder-free gloves
Warning:
To avoid personal contamination with biologically hazardous, toxic, and corrosive
materials, wear chemical-resistant, powder-free gloves when performing this procedure.
To install the UniSpray source:
1. Prepare the instrument for working on the source by stopping solvent flow, ensuring that
the instrument is in Standby mode, and stopping desolvation gas flow.
2. Remove the existing source enclosure.
3. Ensure that the probe assembly is connected to the UniSpray source before you fit the
source to the mass spectrometer.
See also:
4. Use two hands to fit the UniSpray source enclosure to the two supporting studs on the
adapter housing.
Warning:
To avoid harmless, static-like electric shock, ensure that the
instrument is in Standby mode before beginning this procedure.
Warning:
To avoid burn injuries, exercise care when handling the components
of the source enclosure heated to high temperatures. Wait until the hot
components have sufficiently cooled before you handle them.
Installing the UniSpray probe assembly (Page
March 31, 2021, 715004496 Ver. 04 (previously released as Rev. D)
Page 81
201).