See also:
4. If you removed the source enclosure from the device to fit the probe assembly, reinstall the
source enclosure.
See also:
5. Screw the probe assembly's PEEK fitting into the appropriate fluidics port for your
instrument until it is finger tight.
7.29 Replacing the IonSABRE II probe heater
Replace the IonSABRE II probe heater if it fails to heat the probe.
7.29.1 Removing the IonSABRE II probe heater
Required materials
• Chemical-resistant, powder-free gloves
To remove the IonSABRE II probe heater:
Warning:
To avoid personal contamination with biologically hazardous, toxic, and corrosive
materials, wear chemical-resistant, powder-free gloves when performing this procedure.
1. Remove the probe from the source (see
Notice:
To avoid damaging the capillary on instruments where the source is
situated above eye level, remove the source from the device and move it to a
lower position before inserting the probe assembly into the probe.
Removing the UniSpray source (Page
Installing the UniSpray source (Page
March 31, 2021, 715004496 Ver. 04 (previously released as Rev. D)
84).
81).
Removing the IonSABRE II probe (Page
Page 203
51)).