Warning:
To avoid burn injuries, take great care while working with the probe and
source; these components can be hot.
1. Remove the probe from the source (see
2. Retrieve the combined 2.5-mm hex wrench and cone extraction tool from its storage
location on the source adapter housing.
3. Use the combined 2.5-mm hex wrench and cone extraction tool to remove the 3 probe
end-cover retaining screws.
Figure 7–77: Removing the probe end-cover retaining screws
End cover retaining screws
4. Remove the end cover and gasket.
Figure 7–78: Removing the end cover and gasket
Nebulizer adjuster knob
March 31, 2021, 715004496 Ver. 04 (previously released as Rev. D)
Removing the IonSABRE II probe (Page
Page 168
51)).