B.12 Connecting to the electricity source
The mass spectrometer requires a separate, grounded electricity source. The ground connection
in the electrical outlet must be common and connected near the system.
To connect to the electricity source:
Recommendation:
optimum long-term input voltage stability.
Warning:
HAR-type (or better) cords in Europe. The power cords must be replaced only with
ones of adequate rating. For information regarding which cord to use in other countries,
contact your local Waters distributor.
1. Connect the female end of the power cord to the receptacle on the rear panel of the mass
spectrometer.
2. Connect the male end of the mass spectrometer power cord to a suitable 200 to 240 V AC
wall outlet.
B.13 Connecting the camera for the NanoLockSpray source
or ionKey source
The camera for both the NanoLockSpray and ionKey source is connected to the workstation PC
by means of a USB converter.
Figure B–20: Camera connection
ExternalConnections 1
10
9
8
7
6
5
OUT
-
IN
External C onnections 2
10
9
8
7
6
5
OUT -
OUT
OUT
Video output connector
USB converter interface box
Workstation
Use a line conditioner or an uninterruptible power supply (UPS) for
To avoid electric shock, use SVT-type power cords in the United States and
Instrument LAN
4
3
2
1
Service Bus
EPC Com Port
-
I N
EPC Parallel Port
4
3
2
1
Video Output
OUT
10
1
External Connections 1
10
1
External Connections 2
System Activity
March 31, 2021, 715004496 Ver. 04 (previously released as Rev. D)
Page 280