8.3

Electromagnetic compatibility tables

8.3.1

Annex A: Table 1

Guidance and manufacturer's declaration – electromagnetic emissions
The Morpheus is intended for use in the electromagnetic environment specified below. The customer or the
user of the Morpheus should assure that it is used in such an environment.
Emissions test
RF emissions
Group 1
CISPR 11
RF emissions
Class A
CISPR 11
Harmonic emissions
Class A
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
Compliance
IEC 61000-3-3
Morpheus_ND
Compliance
Compliance
Compliance
Class A, B, C, D, o
NOT APPLICABLE
Compliance
Electromagnetic environment –
Virdict
The Morpheus uses RF energy only for its
internal function. Therefore, its RF
emissions are very low and are not likely to
cause any interference in nearby electronic
equipment.
The Morpheus is suitable for use in all
establishments other than domestic, and
may be used in domestic establishments
and those directly connected to the public
low-voltage power supply network that
supplies buildings used for domestic
purposes, provided the following warning is
heeded:
Warning: This equipment/system is
intended for use by healthcare
professionals only. This equipment/ system
may cause radio interference or may
disrupt the operation of nearby equipment.
It may be necessary to take mitigation
measures, such as re-orienting or
relocating the Morpheus or shielding the
location.
guidance
8-17