Connection To Other Equipments - Siare Morpheus ND User Manual

Anaesthesia unit
Table of Contents

Advertisement

4.6

Connection to other equipments

Connection to Siare equipment
If the equipment to be connected is a SIARE unit, all the instructions necessary for
the connection to anaesthesia unit can be found in the documentation supplied with
the unit.
WARNING !! Risk for Patient / User injury
Do not connect external devices NOT manufactured or NOT authorized by SIARE to
the anaesthesia unit (e.g., scavenging systems, patient simulators), and not
described in the present user's manual. In case of need contact SIARE.
WARNING !! Risk of injuries for th patient
When using additional components in the respiratory systems or configurations not
conform to those supplied with the anaesthesia unit, the inspiratory and expiratory
resistence can increase until overcome the standard requirements.
When using configurations of this type, it is necessary to pay particular warning! to
the measuring values.
WARNING !! Risk of electric shock
Morpheus_ND
In case of malfunctioning of ground conductor, the connection of other electric
equipment to the supplementary outlets of the anaesthesia unit could cause an
increase in leakage current over the values allowed by the law.
In case of connection of other devices to the supplementary outlets, it is
necessary to verify the total leakage current.
In case of overcoming of the values allowed for total leakage current, do not
connnect the devices to the supplementary outlets of anaesthesia unit, but to
separate electric outlets.
All the system must comply with the requirements for electromedical equipment
stated by the IEC/EN 60601-1-1 and IEC/EN 60601-1-2 directives.
4-19