Because the mode 3 echo blank limits minimum measurable thicknesses, set the
controls to as short a period as possible; typically no longer than a few hundred
nanoseconds. Special cases may be an exception where mode conversion effects on
curved samples cause significant spurious signals between legitimate echo peaks. In
such cases, set the mode 3 echo blank to as long a period as necessary in order to
prevent detection of the spurious signals.

10.9 Saving Setup Parameters

After adjusting selected waveform parameters, it is possible to store the settings for
quick and easy recall. The 45MG can store up to 35 custom setups in its internal
memory.
To save setup parameters
1.
Make the appropriate changes to the waveform parameters.
2.
Press [2nd F], [FREEZE] (XDCR RECALL).
In the menu, ensure that ACTIVE is highlighted. The changes you made are
located in the active setup.
3.
Press [SAVE].
4.
In the SAVE SETUP screen (see Figure 10-17 on page 184):
a) In the SAVE AS dialog box, enter a name for the custom setup.
b) In the SAVE TO list, select one of the 35 available custom setup locations.
Setups named SE-USER-1 through SE-USER-35 are empty locations. Use caution
when selecting a location that already contains a custom setup, because the new setup
will replace the old information.
c)
Select SAVE to save the custom setup.
CAUTION
Custom Setups for Single Element Transducers
DMTA-10022-01EN, Rev. C, January 2015
183