Figure 10-1 The Active Screen For The Setup Of A Single Element Transducer - Olympus 45MG User Manual

Ultrasonic thickness gage
Hide thumbs Also See for 45MG:
Table of Contents

Advertisement

DMTA-10022-01EN, Rev. C, January 2015
Additionally, when establishing a custom setup for a specific application, it is
essential to verify the performance on reference standards representing the materials
and thickness range to be measured.
To create a custom setup for a single element transducer
1.
Connect the single element transducer to the 45MG (see "Setting Up the
Transducer" on page 51).
2.
Press [2nd F], [FREEZE] (XDCR RECALL).
3.
In the menu, select the CUSTOM SINGLE ELEMENT.
4.
In the CUSTOM SINGLE ELEMENT screen, select the desired custom setup
location (SE-USER-n) in which you want to save the custom setup.
TIP
To reduce the number of changes needing to be made to parameter values, you can
also select an existing single element setup with parameter values that are close to the
desired setup.
5.
In the ACTIVE screen (see Figure 10-1 on page 166):
a) Set DET MODE to the desired detection mode (see "Detection Modes" on
page 169 for details).

Figure 10-1 The ACTIVE screen for the setup of a single element transducer

166
Chapter 10