Spectrum Doppler Low Sensitivity - GE LOGIQ P3 PRO Service Manual

Table of Contents

Advertisement

GE
D
5308917-100, R
IRECTION
7-6-1-6

Spectrum Doppler Low Sensitivity

Spectrum Doppler sensitivity is low.
Is the sensitivity normal for
Is the sensitivity low for all
Is Conn board power
supply on TX64 ok?
Replace the TX64 board.
7-16
PROPRIETARY TO GE
8
EVISION
B Mode?
YES
types of probe?
YES
NO
Figure 7-13 Spectrum Doppler Low Sensitivity
Section 7-6 - Troubleshooting
Note: Run Service Diagnostic test, Refer sec 7-7
for diagnostic test
NO
Go to B Mode Low Sensitivity
section.
NO
Check the probes.
YES
Check the RX64 board.
LOGIQ P3 S
M
ERVICE
ANUAL