3. To reselect continuous operation, press the "Pulse" key
display on the status bar.

5.2.5. Low Dose Imaging

Lose dose imaging is to reduce the exposure dose to the patient.
To produce a Low Dose image
1. Press the "Low Dose" key
2. Press the related X-Ray switch on the foot/handswitch to make low dose fluoro exposures.

5.2.6. Imaging Mode List

The following list summarizes the modes available on the system. Additionally it shows the X-Ray
techniques available with each mode, the status bar and C-Arm control panel display for each mode, and a
brief list of examples for each mode. The examples are not exhaustive and other uses may be as valid as
the examples shown. The user must be aware of the techniques associated with the various modes and
monitor dose rates and doses for each patient.
Auto: X-ray techniques (kVp/mA) are adjusted automatically by system to optimize the image.
Manual: X-ray techniques (kVp/mA) are adjusted manually by user to get the desired image.
The user can select Auto or Manual mode listed in the table below as she/he wants.
Mode
Normal Fluoro
Auto/ Manual
High Level Fluoro
Auto/ Manual
Low Dose
Auto/ Manual
Low Dose High Level
Auto/ Manual
on the C-Arm control panel to enable low dose imaging.
X-Ray
Techniques
kVp:40 to 110
mA: 0.1 to 4
kVp:40 to 110
mA: 0.2 to 12
kVp:40 to 110
mA: 0.1 to 2
kVp:40 to 110
mA: 0.1 to 6
Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product
again. The pulse rates will no longer
Display
General Fluoro Imaging
High quality but high-dose
Imaging
Continuous fluoro imaging
with a lower dose
High quality with a lower
dose
+
Imaging Task
Page 5-11