Power Block - Logiq P3; Tools; Needed Manpower; Preparations - GE LOGIQ P3 PRO Service Manual

Table of Contents

Advertisement

GE
D
5308917-100, R
IRECTION
Section 8-7 Power Block - LOGIQ P3
8-7-1
Transformer Assembly (FRU No. 5408517) Cable Guard Clamp (FRU
No:5219747)
This is a description on how to remove and replace the Transformer Assy.
8-7-1-1

Tools

Common Phillips screwdrivers
8-7-1-2

Needed Manpower

1 persons, 15 minutes
8-7-1-3

Preparations

Shutdown the system and switch off the main Circuit Breaker at the bottom rear side of the system.
NOTE:
For 110V System Configuration use 110V wired Plug (with 3 Amps Fuse).
NOTE:
For 230V System Configuration use 230V Wired plug (With 2 Amps Fuse).
NOTE:
To replace Cable Guard Clamp Donot remove the transformer,remove power cord and press
the cable clamp from both the sides the clamp will come out side.
8-7-1-4

Removal Procedure

1) Remove rear cover. Refer
2) Unscrew two screws (a,b) as marked in
3) Remove GND cable (which ever connected near to "a" ). Refer
8-76
PROPRIETARY TO GE
8
EVISION
section 8-4-3 on page 35
GND
cable
Figure 8-83 Removing Transformer Assembly
Section 8-7 - Power Block - LOGIQ P3
Figure 8-83 on page 8-76
Figure 8-83 on page 8-76
a
b
LOGIQ P3 S